For the CRT Omnidialysis Machine, the errors: ❗1. Outlet Temperature Failure ❗2. Temperature Sensor Failure are both related to dialysate temperature monitoring and control. --- 🔍 WHAT THESE ERRORS MEAN: Error Description Outlet Temperature Failure The temperature of dialysate at the outlet is outside the safe range (usually 41°C or 33°C or it's not being detected correctly. Temperature Sensor Failure The machine is not receiving correct data from the temperature sensor — due to a faulty sensor, broken cable, or internal error. --- ✅ HOW TO TROUBLESHOOT AND FIX: 🔧 Step-by-Step Actions: 1. Restart the Machine Power OFF → wait 1–2 mins → Power ON again. Sometimes errors clear after a reset. --- 2. Check the Dialysate Supply Temperature Ensure incoming RO water is within recommended temperature typically 15–25°C Water that's too cold or hot can trigger this. --- 3. Inspect the Temperature Sensor Probe Locate the outlet temp sensor (usually metal, placed at dialysate output line). Check for: Loose connections Signs of corrosion or water ingress Bent or damaged wires --- 4. Replace or Reseat Sensor If sensor looks damaged → Replace it. If not → unplug and reseat connector tightly. --- 5. Check for Scale or Blockage Scaling or biofilm buildup near the temperature sensor can interfere with readings. Clean the line if needed biomed tech may assist --- 6. Check Internal Electronics If error persists even with a good sensor, the issue may be with the: Sensor signal board Internal wiring Main control board This will need a biomedical technician or CRT service engineer. --- 🧪 Temperature Range Safety Tip: Normal dialysate temp is around 36.5–37.5°C If it goes out of range, machine automatically stops to protect the patient. --- 📞 If Not Solved: Contact CRT Omnidialysis Technical Support or your local service team for: Sensor part replacement Internal diagnostics --- Would you like a template email or report to send to your maintenance/service provider?